GEM30I series

30W AC-DC High-Reliability Interchangeable Medical Adaptor

Features

 Interchangeable AC plugs(plug kit sold separately)
 Medical safety approved(2 x MOPP between primary to secondary)
 Suitable for BF application with appropriate system consideration
 Low leakage current < 100A
 No load power consumption < 0.1W
 Energy efficiency level VI
 Comply with EISA 2007/DoE and EU ErP
 Class 2 power(without earth pin)
 Protections: Short circuit / Overload / Over voltage
 Fully enclosed plastic case
 Optional lock type DC plug
 100% full load burn-in test
 3-year warranty

GEM30I series

ORDER NO.GEM30I05-P1JGEM30I07-P1JGEM30I09-P1JGEM30I12-P1J

OUTPUT

SAFETY MODEL NO.GEM30I05GEM30I07GEM30I09GEM30I12
DC VOLTAGE5V7.5V9V12V
RATED CURRENT4A3.33A3.33A2.5A
CURRENT RANGE0 ~ 4A 0 ~ 3.33A0 ~ 3.33A0 ~ 2.5A
RATED POWER (max.)20W25W30W30W
RIPPLE & NOISE (max.) 100mVp-p 100mVp-p 100mVp-p 100mVp-p
VOLTAGE TOLERANCE ±5.0%±5.0% ±5.0% ±3.0%
LINE REGULATION±1.0%±1.0%±1.0%±1.0%
LOAD REGULATION±5.0%±5.0%±5.0%±3.0%
SETUP, RISE TIME, HOLD UP TIME (Typ.)500ms, 100ms, 16ms/230VAC 500ms, 100ms, 16ms/115VAC at full load

INPUT

VOLTAGE RANGE 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz
EFFICIENCY (Typ.)82% 86% 87% 87%
AC CURRENT (Typ.)1A/ 115VAC 0.6A/ 230VAC
INRUSH CURRENT (Typ.)30A/ 115VAC 65A/ 230VAC
LEAKAGE CURRENT(max.)Touch current < 100 A/264VAC
Specsuntitled-3Download PDF

GEM30I series

ORDER NO.GEM30I15-P1JGEM30I18-P1JGEM30I24-P1JGEM30I48-P1J

OUTPUT

SAFETY MODEL NO.GEM30I15GEM30I18GEM30I24GEM30I48
DC VOLTAGE15V18V24V48V
RATED CURRENT2A1.66A 1.25A 0.625A
CURRENT RANGE0 ~ 2A0 ~ 1.66A0 ~ 1.25A0 ~ 0.625A
RATED POWER (max.)30W30W30W30W
RIPPLE & NOISE (max.) 100mVp-p 100mVp-p 100mVp-p 100mVp-p
VOLTAGE TOLERANCE ±3.0%±2.0%±2.0%±2.0%
LINE REGULATION±1.0%±1.0%±1.0%±1.0%
LOAD REGULATION±3.0%±2.0%±2.0%±2.0%
SETUP, RISE TIME, HOLD UP TIME (Typ.)500ms, 100ms, 16ms/230VAC 500ms, 100ms, 16ms/115VAC at full load

INPUT

VOLTAGE RANGE 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz
EFFICIENCY (Typ.)87% 88%88.5%90%
AC CURRENT (Typ.)1A/ 115VAC 0.6A/ 230VAC
INRUSH CURRENT (Typ.)30A/ 115VAC 65A/ 230VAC
LEAKAGE CURRENT(max.)Touch current < 100 A/264VAC
Specsuntitled-3Download PDF