MPS-30 series

30W Single Output for Medical Type

 

Features

 Universal AC input / Full range
 Low leakage current < 200A
 Protections: Short circuit/ Overload / Over Voltage / Over temperature
 Medical safety approved(2 x MOPP between primary to secondary)
 Fixed switching frequency at 100KHz
 Low cost
 High reliability
 3-year warranty

MPS-30 series

MODELMPS-30-5 MPS-30-12 MPS-30-15MPS-30-24 MPS-30-27MPS-30-48

OUTPUT

DC VOLTAGE5V12V15V24V27V48V
RATED CURRENT5A2.5A2A1.2A 1.1A 0.6A
CURRENT RANGE0 ~ 5A 0 ~ 2.5A 0 ~ 2A 0 ~ 1.2A0 ~ 1.1A 0 ~ 0.6A
RATED POWER25W30W30W28.8W 29.7W28.8W
RIPPLE & NOISE (max.) 80mVp-p120mVp-p 150mVp-p 240mVp-p240mVp-p240mVp-p
VOLTAGE TOLERANCE±3.0%±3.0%±3.0%±3.0%±3.0%± 2.0%
LINE REGULATION±1.0%±1.0%±0.5%±0.5%±0.5%±0.5%
LOAD REGULATION±2.0% ± 1.0%± 1.0%± 1.0%± 1.0%± 1.0%
SETUP, RISE TIME500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)70ms/230VAC 12ms/115VAC at full load

INPUT

VOLTAGE RANGE88 ~ 264VAC auto switch 120 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz
EFFICIENCY(Typ.)72% 75% 76% 77%78% 78%
AC CURRENT (Typ.)0.8A/115VAC 0.5A/230VAC
INRUSH CURRENT (Typ.)COLD START 30A/230VAC
LEAKAGE CURRENT(max.) Earth leakage current < 200 A/264VAC
Specs Sheetuntitled-3Download PDF