MPT-45A series
45W Triple Output Medical Type
Features
Universal AC input / Full range
Low leakage current < 250A
Protections: Short circuit/ Overload / Over Voltage
Low leakage current < 250A
Protections: Short circuit/ Overload / Over Voltage
Cooling by free air convection
Medical safety approved(2 x MOPP between primary to secondary)
100% full load burn-in test
Fixed switching frequency at 45KHz
3-year warranty
Medical safety approved(2 x MOPP between primary to secondary)
100% full load burn-in test
Fixed switching frequency at 45KHz
3-year warranty
MPT-45A series
MODEL | MPT-45A | |||
---|---|---|---|---|
OUTPUT | OUTPUT NUMBER | CH1 | CH2 | CH3 |
OUTPUT VOLTAGE | 5V | 12V | -5V | |
RATED CURRENT | 3A | 2A | 0.3A | |
CURRENT RANGE | 0.4 ~ 5A | 0.2 ~ 2.5A | 0 ~ 0.5A | |
RATED POWER | 40.5W | |||
PEAK LOAD(10sec.) | 52W with 18CFM min. Forced air convection | |||
RIPPLE & NOISE (max.) | 60mVp-p | 120mVp-p | 60mVp-p | |
VOLTAGE TOLERANCE | ±4.0% | ±7.0 | ±5.0% | |
LINE REGULATION | ±1.0% | ±2.0% | ±1.0% | |
LOAD REGULATION | ±3.0% | ±4.0% | ±1.0% | |
SETUP, RISE TIME | 800ms, 20ms/230VAC 800ms, 20ms/115VAC at full load | |||
HOLD UP TIME (Typ.) | 80ms/230VAC 12ms/115VAC at full load | |||
INPUT | VOLTAGE RANGE | 90 ~ 264VAC 127 ~ 370VDC | ||
FREQUENCY RANGE | 47 ~ 63Hz | |||
EFFICIENCY(Typ.) | 73% | |||
AC CURRENT (Typ.) | 1.2A/115VAC 0.7A/230VAC | |||
INRUSH CURRENT (Typ.) | COLD START 20A/115VAC 40A/230VAC | |||
LEAKAGE CURRENT | Earth leakage current < 250 A/264VAC , Touch current < 60 A/264VAC | |||
Specs Sheet | Download PDF |