RPT-160B series
160W Triple Output Medical Type
Features
Universal AC input / Full range
Built-in active PFC function
Protections: Short circuit/ Overload / Over Voltage / Over temperature
5″x3″ compact size
Free air convection for 100W and 145W with 20.5 CFM forced air
Medical safety approved(2 x MOPP between primary to secondary)
With power good and fail signal output
No load power consumption under 0.75W by PS-ON control(G model)
Standby 5V @ 0.8A with fan @ 0.6A without fan(G model)
Suitable for BF application with appropriate system consideration
3-year warranty
RPT-160B series
MODEL | RPT -160B | |||
---|---|---|---|---|
OUTPUT | OUTPUT NUMBER | CH1 | CH2 | CH3 |
DC VOLTAGE | 5V | 12V | -12V | |
RATED CURRENT | 14A | 5A | 1A | |
CURRENT RANGE (convection) | 0.6 ~ 9A | 0.2 ~ 3.4A | 0.1 ~ 0.8A | |
CURRENT RANGE (20.5CFM) | 0.6 ~ 14A | 0.2 ~ 5A | 0.1 ~ 1A | |
RATED POWER(convection) | 98.4W | |||
RATED POWER(20.5CFM) | 146W | |||
RIPPLE & NOISE (max.) | 100mVp-p | 120mVp-p | 120mVp-p | |
VOLTAGE ADJ. RANGE | CH1:5 ~ 5.5V | |||
VOLTAGE TOLERANCE | ±2.0% | ±5.0% | -4,+5% | |
LINE REGULATION | ±0.5% | ±1.0% | ±1.0% | |
LOAD REGULATION | ±1.5% | ±3.0% | -4,+5% | |
SETUP, RISE TIME | 1800ms, 30ms/230VAC 3500ms, 30ms/115VAC at full load | |||
HOLD UP TIME (Typ.) | 16ms/230VAC/115VAC at full load | |||
INPUT | VOLTAGE RANGE | 90 ~ 264VAC 127 ~ 370VDC | ||
FREQUENCY RANGE | 47 ~ 63Hz | |||
POWER FACTOR (Typ.) | PF>0.93/230VAC PF>0.98/115VAC at full load | |||
EFFICIENCY(Typ.) | 84% | |||
AC CURRENT (Typ.) | 1.8A/115VAC 0.9A/230VAC | |||
INRUSH CURRENT (Typ.) | COLD START 35A/115VAC 70A/230VAC | |||
LEAKAGE CURRENT | Earth leakage current < 200 A/264VAC , Touch current < 100 A/264VAC | |||
Specs Sheet | Download PDF |