GSM25B series
25W AC-DC High-Reliability Medical Adaptor
Features
Universal AC input / Full range
2 pole AC inlet IEC320-C8
Medical safety approved(2 x MOPP between primary to secondary)
Suitable for BF application with appropriate system consideration
Low leakage current < 50A
No load power consumption < 0.1W
Energy efficiency level VI(except 5~9V for level V)
Comply with EISA 2007/DoE, NRCan, AU/NZ MEPS, EU ErP and meet CoC Version 5
High efficiency up to 88%
High operating temperature up to +60C
Class 2 power(without earth pin)
Protections: Short circuit / Overload / Over voltage
Fully enclosed plastic case
LED indicator for power on
Optional lock type DC plug
100% full load burn-in test
3-year warranty
GSM25B series
ORDER NO. | GSM25B05-P1J | GSM25B07-P1J | GSM25B09-P1J | GSM25B12-P1J | |
---|---|---|---|---|---|
OUTPUT | SAFETY MODEL NO. | GSM25B05 | GSM25B07 | GSM25B09 | GSM25B12 |
DC VOLTAGE | 5V | 7.5V | 9V | 12V | |
RATED CURRENT | 4A | 2.93A | 2.77A | 2.08A | |
CURRENT RANGE | 0 ~ 4A | 0 ~ 2.93A | 0 ~ 2.77A | 0 ~ 2.08A | |
RATED POWER (max.) | 20W | 22W | 25W | 25W | |
RIPPLE & NOISE (max.) | 60mVp-p | 80mVp-p | 80mVp-p | 120mVp-p | |
VOLTAGE TOLERANCE | ±6.0% | ±5.0% | ±5.0% | ±3.0% | |
LINE REGULATION | ±1.0% | ±1.0% | ±1.0% | ±1.0% | |
LOAD REGULATION | ±6.0% | ±5.0% | ±5.0% | ±3.0% | |
SETUP, RISE TIME | 500ms, 30ms / 230VAC 1000ms, 30ms / 115VAC at full load | ||||
HOLD UP TIME (Typ.) | 16ms / 230VAC 16ms / 115VAC at full load | ||||
INPUT | VOLTAGE RANGE | 80 ~ 264VAC 113 ~ 370VDC | |||
FREQUENCY RANGE | 47 ~ 63Hz | 47 ~ 63Hz | 47 ~ 63Hz | 47 ~ 63Hz | |
EFFICIENCY (Typ.) | 80% | 83% | 84% | 86% | |
AC CURRENT (Typ.) | 0.7A/ 115VAC 0.35A/ 230VAC | ||||
INRUSH CURRENT (Typ.) | 55A/ 230VAC 30A/ 115VAC | ||||
LEAKAGE CURRENT(max.) | Touch current < 50 A/264VAC | ||||
Specs | ![]() |
GSM25B series
ORDER NO. | GSM25B15-P1J | GSM25B18-P1J | GSM25B24-P1J-P1J | GSM25B48-P1J | |
---|---|---|---|---|---|
OUTPUT | SAFETY MODEL NO. | GSM25B15 | GSM25B18 | GSM25B24 | GSM25B48 |
DC VOLTAGE | 5V | 7.5V | 9V | 12V | |
RATED CURRENT | 1.66A | 1.38A | 1.04A | 0.52A | |
CURRENT RANGE | 0 ~ 1.66A | 0 ~ 1.38A | 0 ~ 1.04A | 0 ~ 0.52A | |
RATED POWER (max.) | 25W | 25W | 25W | 25W | |
RIPPLE & NOISE (max.) | 150mVp-p | 150mVp-p | 180mVp-p | 240mVp-p | |
VOLTAGE TOLERANCE | ±3.0% | ±3.0% | ±2.0% | ±2.0% | |
LINE REGULATION | ±1.0% | ±1.0% | ±1.0% | ±1.0% | |
LOAD REGULATION | ±3.0% | ±3.0% | ±2.0% | ±2.0% | |
SETUP, RISE TIME | 500ms, 30ms / 230VAC 1000ms, 30ms / 115VAC at full load | ||||
HOLD UP TIME (Typ.) | 16ms / 230VAC 16ms / 115VAC at full load | ||||
INPUT | VOLTAGE RANGE | 80 ~ 264VAC 113 ~ 370VDC | |||
FREQUENCY RANGE | 47 ~ 63Hz | 47 ~ 63Hz | 47 ~ 63Hz | 47 ~ 63Hz | |
EFFICIENCY (Typ.) | 86% | 86% | 87% | 88% | |
AC CURRENT (Typ.) | 0.7A/ 115VAC 0.35A/ 230VAC | ||||
INRUSH CURRENT (Typ.) | 55A/ 230VAC 30A/ 115VAC | ||||
LEAKAGE CURRENT(max.) | Touch current < 50 A/264VAC | ||||
Specs | ![]() |