RPS-60 series

65W Single Output Medical Type

 

Features

 4″x2″ miniature size
 Universal AC input / Full range
 Low leakage current < 200A
 Protections: Short circuit/ Overload / Over Voltage
 Cooling by free air convection
 UL60601-1/IEC60601-1/EN60601-1 medical safety approved
 UL60950-1/IEC60950-1/EN60950-1 ITE safety approved
 No load power consumption < 0.75W
 Fixed switch frequency at 100KHz

RPS-60 series

MODELRPS-60-3.3RPS-60-5 RPS-60-12

OUTPUT

DC VOLTAGE3.3V 5V12V
RATED CURRENT10A10A5A
CURRENT RANGE0 ~ 11A0 ~ 11A0 ~ 5.5A
RATED POWER33W50W60W
PEAK LOAD(10sec.)36.3W55W66W
RIPPLE & NOISE (max.) 80mVp-p80mVp-p120mVp-p
VOLTAGE ADJ. RANGE3.1 ~ 3.6V 4.75 ~ 5.5V11.4 ~ 13.2V
VOLTAGE TOLERANCE±2.0%±2.0%±2.0%
LINE REGULATION±0.5%±0.5%±0.5%
LOAD REGULATION±1.0%±1.0%±1.0%
SETUP, RISE TIME500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)50ms/230VAC 13ms/115VAC at full load

INPUT

VOLTAGE RANGE90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz
EFFICIENCY(Typ.)74%79% 83%
AC CURRENT (Typ.)1.8A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)COLD START 60A/230VAC 30A/115VAC
LEAKAGE CURRENT(max.) For earth <200uA/ 264VAC, For patient <100uA/264VAC
Specs Sheetuntitled-3Download PDF

RPS-60 series

MODELRPS-65-15RPS-65-24RPS-65-48

OUTPUT

DC VOLTAGE15V24V48V
RATED CURRENT4A2.5A1.25A
CURRENT RANGE0 ~ 4.4A0 ~ 2.75A0 ~ 1.375A
RATED POWER60W60W60W
PEAK LOAD(10sec.)66W66W66W
RIPPLE & NOISE (max.) 120mVp-p 240mVp-p 300mVp-p
VOLTAGE ADJ. RANGE13.5 ~ 16.5V22.8 ~ 27.6V 45.6 ~ 52.8V
VOLTAGE TOLERANCE±2.0%±1.0%±1.0%
LINE REGULATION±0.5%±0.5%±0.5%
LOAD REGULATION±1.0%±1.0%±1.0%
SETUP, RISE TIME500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)50ms/230VAC 13ms/115VAC at full load

INPUT

VOLTAGE RANGE90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz
EFFICIENCY(Typ.)84%85% 86%
AC CURRENT (Typ.)1.8A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)COLD START 60A/230VAC 30A/115VAC
LEAKAGE CURRENT(max.) For earth <200uA/ 264VAC, For patient <100uA/264VAC
Specs Sheetuntitled-3Download PDF