RPD-60B series

60W Reliable Dual Output Medical Grade
Features
4″x2″ compact size
Medical safety approved(2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1
Suitable for BF application with appropriate system consideration
Cooling by free air convection
EMI class B for class 1 configuration
Extremely low leakage current
Protections: Short circuit/ Overload / Over Voltage
3-year warranty
RPD-60B series
| MODEL | RPD-60B | ||
|---|---|---|---|
OUTPUT | OUTPUT NUMBER | CH1 | CH2 |
| OUTPUT VOLTAGE | 5V | 24V | |
| RATED CURRENT | 3.5A | 1.5A | |
| CURRENT RANGE | 0.5 ~ 3.85A | 0.1 ~ 1.65A | |
| RATED POWER | 53.5W | ||
| PEAK LOAD(10sec.) | 58.85W | ||
| RIPPLE & NOISE (max.) | 80mVp-p | 100mVp-p | |
| VOLTAGE TOLERANCE | +3,-2% | +8,-4% | |
| LINE REGULATION | ±0.5% | ±1.0% | |
| LOAD REGULATION | ±1.5% | ±2.0% | |
| SETUP, RISE TIME | 300ms, 15ms/230VAC 300ms, 15ms/115VAC at full load | ||
| HOLD UP TIME (Typ.) | 70ms/230VAC 14ms/115VAC at full load | ||
INPUT | VOLTAGE RANGE | 90 ~ 264VAC 127 ~ 370VDC | |
| FREQUENCY RANGE | 47 ~ 63Hz | ||
| EFFICIENCY(Typ.) | 82% | ||
| AC CURRENT (Typ.) | 1.1A/115VAC 0.7A/230VAC | ||
| INRUSH CURRENT (Typ.) | COLD START 60A/230VAC 30A/115VAC | ||
| LEAKAGE CURRENT | Earth leakage current < 150 A/264VAC , Touch current < 100 A/264VAC | ||
| Specs Sheet | |||