300W Reliable Green Medical Power Supply
Features
5″x3″ compact size
Medical safety approved(2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1
Suitable for BF application with appropriate system consideration
200W convection, 300W force air
EMI Class B for both Class 1 & Class A for Class 2 configuration
No load power consumption <0.5W by PS-ON control
Extremely low leakage current
5Vdc standby output, 12Vdc fan supply, Power Good, Power Fail and remote sense
Protections: Short circuit/ Overload / Over Voltage / Over temperature
Typical Lifetime > 40K hours
3-year warranty
RPS-300 series
MODEL | RPS-300- 12 | RPS-300- 15 | RPS-300- 24 | RPS-300- 27 | RPS-300- 48 |
||
---|---|---|---|---|---|---|---|
OUTPUT | DC VOLTAGE | 12V | 15V | 24V | 27V | 48V | |
RATED CURRENT (20.5CFM) | 25A | 20A | 12.5A | 11.12A | 6.25A | ||
CURRENT | Convection | 0 ~ 16.67A | 0 ~ 13.33A | 0 ~ 8.33A | 0 ~ 7.4A | 0 ~ 4.17A | |
20.5CFM | 0 ~ 25A | 0 ~ 20A | 0 ~ 12.5A | 0 ~ 11.12A | 0 ~ 6.25A | ||
RATED POWER | Convection | 200W | 200W | 200W | 200W | 200.2W | |
20.5CFM | 300W | 300W | 300W | 300W | 300W | ||
RIPPLE & NOISE (max.) | 120mVp-p | 120mVp-p | 150mVp-p | 200mVp-p | 250mVp-p | ||
VOLTAGE ADJ. RANGE | 11.4 ~ 12.6V | 14.25 ~ 15.75V | 22.8 ~ 25.2V | 25.65 ~ 28.35V | 45.6 ~ 50.4V | ||
VOLTAGE TOLERANCE | ±3.0% | ±3.0% | ±2.0% | ±2.0% | ±2.0% | ||
LINE REGULATION | ±0.5% | ±0.5% | ±0.5% | ±0.5% | ±0.5% | ||
LOAD REGULATION | ±1.0% | ±1.0% | ±1.0% | ±1.0% | ±1.0% | ||
SETUP, RISE TIME | 2500ms, 30ms/230VAC 3000ms, 30ms/115VAC at full load | ||||||
HOLD UP TIME (Typ.) | 16ms/230VAC/115VAC at full load | ||||||
INPUT | VOLTAGE RANGE | 90 ~ 264VAC 127 ~ 370VDC | |||||
FREQUENCY RANGE | 47 ~ 63Hz | ||||||
POWER FACTOR (Typ.) | PF>0.93/230VAC PF>0.98/115VAC at full load | ||||||
EFFICIENCY(Typ.) | 90% | 90% | 92.5% | 93% | 93% | ||
AC CURRENT (Typ.) | 3.5A/115VAC 1.8A/230VAC | ||||||
INRUSH CURRENT (Typ.) | COLD START 35A/115VAC 70A/230VAC | ||||||
LEAKAGE CURRENT(max.) | PCB Type: Earth leakage current <150 A/ 264VAC, Touch current <70 A/264VAC Enclosed Type: Earth leakage current <200 A/ 264VAC, Touch current <70 A/264VAC |
||||||
Specs Sheet | Download PDF |