GSM90B series

90W AC-DC Reliability Green Medical Adaptor

Features

 2 pole AC inlet IEC320-C8, Class 2 power unit
 Medical safety approved(2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1
 Extremely Low leakage current
 No load power consumption < 0.15W
 Energy efficiency level VI and meet CoC version 5
 -30~70C wide range working temperature
 Protections: Short circuit / Overload / Over voltage/ Over temperature
 LED indicator for power on
 Lifetime > 90 K hours
 3-year warranty

GSM90B series

ORDER NO.GSM90B12-P1M GSM90B15-P1MGSM90B19-P1M GSM90B24-P1MGSM90B48-P1M

OUTPUT

SAFETY MODEL NO.GSM90B12GSM90B15GSM90B19GSM90B24GSM90B48
DC VOLTAGE12V15V19V24V48V
RATED CURRENT6.67A6A4.74A 3.75A 1.87A
CURRENT RANGE0 ~ 6.67A 0 ~ 6A 0 ~ 4.74A 0 ~ 3.75A 0 ~ 1.87A
RATED POWER (max.)80W90W90W90W90W
RIPPLE & NOISE (max.) 120mVp-p120mVp-p120mVp-p180mVp-p 200mVp-p
VOLTAGE TOLERANCE ±5.0% ±5.0%±4.0%±3.0% ±2.5%
LINE REGULATION±1.0%±1.0%±1.0%±1.0%±1.0%
LOAD REGULATION±5.0% ±5.0%±4.0%±3.0% ±2.5%
SETUP, RISE TIME 1000ms, 50ms / 230VAC 1500ms, 50ms / 115VAC at full load
HOLD UP TIME (Typ.)30ms / 230VAC 20ms / 115VAC at full load

INPUT

VOLTAGE RANGE 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz
POWER FACTOR (Typ.)PF>0.91 / 230VAC PF>0.95 / 115VAC at full load
EFFICIENCY (Typ.)88% 89%89%90%91%
AC CURRENT (Typ.)1.3A/ 115VAC 0.6A/ 230VAC
INRUSH CURRENT (Typ.)Cold start 30A/ 115VAC 60A/ 230VAC
LEAKAGE CURRENT(max.)Touch current < 100 A/264VAC
Specsuntitled-3Download PDF