MPT-65B series

65W Triple Output Medical Type

Features

 Universal AC input / Full range
 Low leakage current < 250A
 Protections: Short circuit/ Overload / Over Voltage
 Cooling by free air convection
 Medical safety approved(2 x MOPP between primary to secondary)
 100% full load burn-in test
 Fixed switching frequency at 45KHz
 3-year warranty

MPT-65B series

MODELMPT-65B

OUTPUT

OUTPUT NUMBERCH1CH2CH3
DC VOLTAGE5V12V-12V
RATED CURRENT5.5A 2.5A0.3A
CURRENT RANGE0.4 ~ 7A0.2 ~ 3.2A0 ~ 0.7A
RATED POWER63.5W
OUTPUT POWER (max.)72W with 18CFM min. Forced air convection
RIPPLE & NOISE (max.) 60mVp-p160mVp-p 100mVp-p
VOLTAGE ADJ. RANGECH1:4.5 ~ 5.5V
VOLTAGE TOLERANCE±4.0%+10,-7%±5.0%
LINE REGULATION±1.0% ±2.0%±1.0%
LOAD REGULATION±3.0%±4.0% ±1.0%
SETUP, RISE TIME800ms, 20ms/230VAC 800ms, 20ms/115VAC at full load
HOLD UP TIME (Typ.)80ms/230VAC 12ms/115VAC at full load

INPUT

VOLTAGE RANGE90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz
EFFICIENCY(Typ.)74%
AC CURRENT (Typ.)1.6A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)COLD START 20A/115VAC 40A/230VAC
LEAKAGE CURRENTEarth leakage current < 250 A/264VAC , Touch current < 60 A/264VAC
Specs Sheetuntitled-3Download PDF