MPM-05
5W High Reliable Green Medical Encapsulated Type
Features
1.8″x1″ compact size
Medical safety approved(2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1
Suitable for BF application with appropriate system consideration
No load power consumption < 0.075W
Extremely low leakage current
Wide operation temp. range -40 ~ +85C
EMI class B for class 2 configuration
Protections: Short circuit/ Overload / Over Voltage / Over temperature
No minimum load required
Typical Lifetime > 48K hours
3-year warranty
MPM-05 series
MODEL | MPM-05-3.3 | MPM-05-5 | MPM-05-12 | MPM-05-15 | MPM-05-24 | |
---|---|---|---|---|---|---|
OUTPUT | DC VOLTAGE | 3.3V | 5V | 12V | 15V | 24V |
RATED CURRENT | 1.25A | 1A | 0.42A | 0.33A | 0.23A | |
CURRENT RANGE | 0 ~ 1.25A | 0 ~ 1A | 0 ~ 0.42A | 0 ~ 0.33A | 0 ~ 0.23A | |
PEAK CURRENT | 1.38A | 1.1A | 0.46A | 0.36A | 0.25A | |
RATED POWER | 4.1W | 5W | 5W | 5W | 5.5W | |
PEAK LOAD(10sec.) | 4.6W | 5.5W | 5.5W | 5.4W | 6W | |
RIPPLE & NOISE (max.) | 100mVp-p | 100mVp-p | 150mVp-p | 150mVp-p | 180mVp-p | |
VOLTAGE TOLERANCE | ±2.5% | ±2.5% | ±2.5% | ±2.5% | ±2.5% | |
LINE REGULATION | ±0.3% | ±0.3% | ±0.3% | ±0.3% | ±0.3% | |
LOAD REGULATION | ±0.5% | ±0.5% | ±0.5% | ±0.5% | ±0.5% | |
SETUP, RISE TIME | 1000ms, 30ms/230VAC 1000ms, 30ms/115VAC at full load | |||||
HOLD UP TIME (Typ.) | 40ms/230VAC 12ms/115VAC at full load | |||||
INPUT | VOLTAGE RANGE | 80 ~ 264VAC | ||||
FREQUENCY RANGE | 47 ~ 63Hz | |||||
EFFICIENCY(Typ.) | 74% | 80% | 80% | 81% | 82% | |
AC CURRENT (Typ.) | 0.2A/115VAC 0.1A/230VAC | |||||
INRUSH CURRENT (Typ.) | COLD START 25A/115VAC 45A/230VAC | |||||
LEAKAGE CURRENT(max.) | Touch current <80 A/264VAC | |||||
Specs Sheet | Download PDF |