MPM-30-xST series
30W High Reliable Green Medical Encapsulated Type
Features
• Universal AC input / Full range
• Medical safety (2xMOPP)
• Suitable for BF application with appropriate system consideration
• Protections: Short circuit / Overload / Over voltage / Over temp.
• Extremely low leakage current
• No load power consumption <0.075W
• Isolation Class II
• Medical safety (2xMOPP)
• Suitable for BF application with appropriate system consideration
• Protections: Short circuit / Overload / Over voltage / Over temp.
• Extremely low leakage current
• No load power consumption <0.075W
• Isolation Class II
• -40~+85 o C wide range working temperature
• Withstand 5G vibration test
• Miniature size, high power density
• Fully isolated plastic case
• 3 years warranty
• Withstand 5G vibration test
• Miniature size, high power density
• Fully isolated plastic case
• 3 years warranty
MPM-30-xST series
MODEL No. | MPM-30-xST series |
---|---|
AC input voltage range | 80~264VAC |
Leakage current | <80uA at 264VAC |
AC inrush current (max.) | Cold start, 25A at 115VAC, 45A at 230VAC |
Overload protection | 105%~160% hiccup mode, auto-recovery |
Over voltage protection | 105%~135% shut down o/p voltage |
Setup, rise, hole up time | 500ms, 30ms, 40ms at 230VAC |
Withstand voltage | I/P-O/P: 4kVAC |
Working temperature | -40~+85°C (refer to output derating curve) |
Safety standards | IEC60601-1, EN60601-1, EAC TP TC 004, UL ANSI/AAMI ES60601, CAN/CSA-C22 3rd edition; Design refer to 60335-1 |
EMC standards | EN55011 class B, EN61000-3-2,3, EN61000-4-2,3,4,5,6,8,11, EN60601-1-2 |
Connection | 4 pins |
Dimension (LxWxH)(mm) | PCB mounting: 69.5x 39x 24; Screw I/O: 91x 39.5x 28.5 |
Specs | Download PDF |