MPM-15

15W High Reliable Green Medical Encapsulated Type

 

Features

 2.06″x1.07″ compact size
 Medical safety approved(2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1
 Suitable for BF application with appropriate system consideration
 No load power consumption < 0.075W(0.09W for 3.3V)
 Extremely low leakage current
 Wide operation temp. range -40 ~ +85C
 EMI class B for class 2 configuration
 Protections: Short circuit/ Overload / Over Voltage / Over temperature
 No minimum load required
 3-year warranty

MPM-15 series

MODELMPM-15-3.3 MPM-15-5 MPM-15-12 MPM-15-15 MPM-15-24

OUTPUT

DC VOLTAGE3.3V5V12V15V24V
RATED CURRENT3.5A3A 1.25A 1A0.63A
CURRENT RANGE0 ~ 3.5A0 ~ 3A 0 ~ 1.25A0 ~ 1A0 ~ 0.63A
PEAK CURRENT3.85A3.3A 1.38A 1.1A 0.69A
RATED POWER11.6W 15W15W15W15.1W
PEAK LOAD(10sec.)12.7W 16.5W 16.6W16.5W 16.6W
RIPPLE & NOISE (max.) 150mVp-p150mVp-p150mVp-p180mVp-p 180mVp-p
VOLTAGE TOLERANCE±1.5%±1.5%±1.5%±1.5%±1.5%
LINE REGULATION±0.5%±0.5%±0.3%±0.3%±0.3%
LOAD REGULATION±0.5%±0.5%±0.5%±0.5%±0.5%
SETUP, RISE TIME1500ms, 30ms/230VAC 1500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)40ms/230VAC 10ms/115VAC at full load

INPUT

VOLTAGE RANGE80 ~ 264VAC
FREQUENCY RANGE47 ~ 63Hz
EFFICIENCY(Typ.)83.5% 85.5%86.5%87% 86.5%
AC CURRENT (Typ.)0.6A/115VAC 0.3A/230VAC
INRUSH CURRENT (Typ.)COLD START 20A/115VAC 45A/230VAC
LEAKAGE CURRENT(max.) Touch current <80 A/264VAC
Specs Sheetuntitled-3Download PDF