MPM-20

20W High Reliable Green Medical Encapsulated Type

 

Features

 2.06″x1.07″ compact size
 Medical safety approved(2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1
 Suitable for BF application with appropriate system consideration
 No load power consumption < 0.075W(0.09W for 3.3V)
 Extremely low leakage current
 Wide operation temp. range -35 ~ +85C
 EMI class B for class 2 configuration
 Protections: Short circuit/ Overload / Over Voltage / Over temperature
 No minimum load required
 3-year warranty

MPM-20 series

MODELMPM-20-3.3 MPM-20-5 MPM-20-12 MPM-20-15 MPM-20-24

OUTPUT

DC VOLTAGE3.3V5V12V15V24V
RATED CURRENT4.5A4A1.8A1.4A0.9A
CURRENT RANGE0 ~ 4.5A0 ~ 4A0 ~ 1.8A 0 ~ 1.4A 0 ~ 0.9A
PEAK CURRENT4.95A4.4A1.98A1.54A0.99A
RATED POWER14.9W20W21.6W 21W21.6W
PEAK LOAD(10sec.)16.3W22W23.8W23.1W23.8W
RIPPLE & NOISE (max.) 150mVp-p150mVp-p150mVp-p180mVp-p 180mVp-p
VOLTAGE TOLERANCE±1.5%±1.5%±1.5%±1.5%±1.5%
LINE REGULATION±0.5%±0.5%±0.3%±0.3%±0.3%
LOAD REGULATION±0.5%±0.5%±0.5%±0.5%±0.5%
SETUP, RISE TIME1500ms, 30ms/230VAC 1500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)40ms/230VAC 10ms/115VAC at full load

INPUT

VOLTAGE RANGE80 ~ 264VAC
FREQUENCY RANGE47 ~ 63Hz
EFFICIENCY(Typ.)81% 85%85.5%87% 87%
AC CURRENT (Typ.)0.75A/115VAC 0.5A/230VAC
INRUSH CURRENT (Typ.)COLD START 20A/115VAC 45A/230VAC
LEAKAGE CURRENT(max.) Touch current <80 A/264VAC
Specs Sheetuntitled-3Download PDF